Nikon TE2000U  Dual mode / Time lapse / Ratio Imaging  Microscope

Fluorescence Filter Settings for the DG-4
 

 

 

REQUIRES

SimplePCI
Filter Setting

Light at the specimen

Dichroic filtercube position on the Microscope

AND

DG-4 Filter
(excitation plate)

in DG-4 slot #

Default Idle Position

No light

Any

 

n/a

n/a

WHITE LIGHT

White Light to filtercube turret

 **For single-channel filtercubes ONLY**

UV (DAPI)   

 

"Open"


4
 

    or

B (FITC, GFP, AlexaFluor 488) 

    or
G (Texas Red, Rhodamine)

    or

R (Cy5, AlexaFluor 647 or 633)

DAPI

UV light for excitation of DAPI

M* (multiple excitation)

 

DAPI
403/12 nm
or
"Open"

4

1

    or

UV (DAPI)

FITC

Excitation for green-emitting fluorophores

M* (multiple excitation)

 

FITC
492/19 nm or
"Open"

3

1

    or

B (FITC)

Texas Red

Excitation for

red-emitting fluorophores

M* (multiple excitation)

 

TX RD
572/23 nm
or
"Open"

2

1

    or

G (Texas Red, Rhodamine)

380nm

FURA 380nm excitation

F-2

 

380nm

2 (3^)

340nm

FURA 340 excitation

F-2

 

340nm

1 (4^)

manDAPI

White Light to filtercube turret

UV (DAPI)  

"Open"

1

manFITC

White Light to filtercube turret

B (FITC)  

"Open"

1

manRhod

White Light to filtercube turret

G (Texas Red, Rhodamine)  

"Open"

1

man647

White Light to filtercube turret

R** (far-red emission: Cy5, Alexa 647)  

"Open"

1

CFP/YFP 430/500 M* (86002V2   C & YFP)   430/25 & 500/20 3 / 4

PRIM 470/20

Excitation at 470nm for PRIM

Not currently installed.  See Facility Management.

 

PRIM 470/20

4

PRIM 410/40

Excitation at 410nm for PRIM

Not currently installed.  See Facility Management.

 

PRIM 410/40

3

Transmitted 
(Uniblitz)

Shutter for transmitted light

When Transmitted light only: "T" dichroic position.  Also see "Setting up for transmitted light imaging" for configuration of transmitted light path.

 With Fluorescence: Set filters for corresponding fluorescence imaging.
Set Front switches to n.c. and remote

* The dichroic cube in the M position may need changing.  Alternate cubes are stored in the plastic drawer.  Move carousel 2 positions counter clockwise from the M position and carefully pull out the dichroic from the left side access port.  Please store removed cube in one of the plastic storage cubes.

^ These are the old positions and may be what are programmed in your profile.  Imaging is superior with the new settings.

** The filtercube for far-red flruoescence may not be installed in the dichroic turret. 

 